| 64T | ¥1960.00 |
| 112T | ¥2800.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 3.106 | 3.055 |
| 5.00 | 2.024 | 1.973 |
| 2.50 | 1.179 | 1.129 |
| 1.25 | 0.691 | 0.641 |
| 0.63 | 0.372 | 0.322 |
| 0.32 | 0.217 | 0.167 |
| 0.16 | 0.144 | 0.094 |
| 0.00 | 0.050 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-110% | 101% |
| EDTA plasma(n=5) | 88-105% | 95% |
| Heparin plasma(n=5) | 90-102% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-105% | 85-102% | 91-110% | 91-108% |
| EDTA plasma(n=5) | 90-104% | 90-106% | 88-108% | 92-109% |
| Heparin plasma(n=5) | 88-103% | 94-108% | 95-105% | 88-107% |