| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.185 | 2.112 |
| 50.00 | 1.546 | 1.473 |
| 25.00 | 1.064 | 0.991 |
| 12.50 | 0.907 | 0.834 |
| 6.25 | 0.542 | 0.469 |
| 3.13 | 0.291 | 0.218 |
| 1.56 | 0.185 | 0.112 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-116% | 105% |
| EDTA plasma(n=5) | 82-92% | 87% |
| Heparin plasma(n=5) | 86-104% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-105% | 82-102% | 92-120% | 84-90% |
| EDTA plasma(n=5) | 82-100% | 84-115% | 87-112% | 80-116% |
| Heparin plasma(n=5) | 94-118% | 86-107% | 93-113% | 89-99% |

關(guān)注微信公眾號(hào)