| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.075 | 1.967 |
| 50.00 | 1.502 | 1.394 |
| 25.00 | 1.128 | 1.020 |
| 12.50 | 0.928 | 0.820 |
| 6.25 | 0.539 | 0.431 |
| 3.13 | 0.387 | 0.279 |
| 1.57 | 0.226 | 0.118 |
| 0.00 | 0.108 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 80-93% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-99% | 98-106% | 86-99% | 78-95% |
| EDTA plasma(n=5) | 89-99% | 83-99% | 98-105% | 96-101% |
| Heparin plasma(n=5) | 85-100% | 84-99% | 85-97% | 88-102% |