| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.075 | 1.989 |
| 100.00 | 1.466 | 1.380 |
| 50.00 | 1.198 | 1.112 |
| 25.00 | 0.959 | 0.873 |
| 12.50 | 0.533 | 0.447 |
| 6.25 | 0.356 | 0.270 |
| 3.13 | 0.196 | 0.110 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-98% | 93% |
| EDTA plasma(n=5) | 84-107% | 96% |
| Heparin plasma(n=5) | 92-101% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 80-97% | 93-111% | 89-102% | 90-98% |
| EDTA plasma(n=5) | 91-113% | 85-119% | 80-95% | 86-114% |
| Heparin plasma(n=5) | 90-107% | 87-111% | 89-104% | 90-110% |