| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.022 | 1.938 |
| 50.00 | 1.459 | 1.375 |
| 25.00 | 0.895 | 0.811 |
| 12.50 | 0.739 | 0.655 |
| 6.25 | 0.515 | 0.431 |
| 3.13 | 0.345 | 0.261 |
| 1.56 | 0.163 | 0.079 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-113% | 104% |
| EDTA plasma(n=5) | 89-113% | 101% |
| Heparin plasma(n=5) | 89-113% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-98% | 83-93% | 91-118% | 80-109% |
| EDTA plasma(n=5) | 90-117% | 83-112% | 85-109% | 90-111% |
| Heparin plasma(n=5) | 89-97% | 90-102% | 83-98% | 82-108% |