| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.132 | 2.048 |
| 5.00 | 1.605 | 1.521 |
| 2.50 | 1.008 | 0.924 |
| 1.25 | 0.825 | 0.741 |
| 0.63 | 0.628 | 0.544 |
| 0.31 | 0.257 | 0.173 |
| 0.16 | 0.126 | 0.042 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-114% | 104% |
| EDTA plasma(n=5) | 86-92% | 89% |
| Heparin plasma(n=5) | 93-118% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-106% | 85-119% | 89-113% | 85-107% |
| EDTA plasma(n=5) | 84-107% | 88-106% | 86-94% | 89-112% |
| Heparin plasma(n=5) | 91-112% | 92-100% | 86-112% | 85-94% |