| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.156 | 2.069 |
| 500.00 | 1.714 | 1.627 |
| 250.00 | 1.050 | 0.963 |
| 125.00 | 0.680 | 0.593 |
| 62.50 | 0.543 | 0.456 |
| 31.25 | 0.364 | 0.277 |
| 15.63 | 0.190 | 0.103 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-113% | 103% |
| EDTA plasma(n=5) | 93-102% | 98% |
| Heparin plasma(n=5) | 86-113% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-89% | 80-90% | 90-96% | 90-104% |
| EDTA plasma(n=5) | 90-110% | 94-102% | 87-115% | 84-115% |
| Heparin plasma(n=5) | 81-120% | 83-112% | 82-113% | 84-111% |