| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.049 | 1.983 |
| 50.00 | 1.499 | 1.433 |
| 25.00 | 1.078 | 1.012 |
| 12.50 | 0.753 | 0.687 |
| 6.25 | 0.567 | 0.501 |
| 3.13 | 0.351 | 0.285 |
| 1.56 | 0.171 | 0.105 |
| 0.00 | 0.066 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-117% | 104% |
| EDTA plasma(n=5) | 90-111% | 101% |
| Heparin plasma(n=5) | 81-113% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-117% | 86-102% | 90-96% | 85-96% |
| EDTA plasma(n=5) | 93-101% | 86-99% | 88-112% | 95-111% |
| Heparin plasma(n=5) | 84-109% | 84-114% | 81-88% | 82-89% |

關(guān)注微信公眾號