| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.079 | 1.996 |
| 50.00 | 1.620 | 1.537 |
| 25.00 | 1.238 | 1.155 |
| 12.50 | 0.945 | 0.862 |
| 6.25 | 0.626 | 0.543 |
| 3.13 | 0.376 | 0.293 |
| 1.56 | 0.297 | 0.214 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-100% | 96% |
| EDTA plasma(n=5) | 91-126% | 109% |
| Heparin plasma(n=5) | 93-114% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-100% | 86-111% | 94-118% | 86-115% |
| EDTA plasma(n=5) | 88-111% | 86-111% | 82-107% | 93-107% |
| Heparin plasma(n=5) | 87-104% | 82-110% | 84-109% | 81-94% |

關(guān)注微信公眾號(hào)