| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.204 | 2.117 |
| 25.00 | 1.560 | 1.473 |
| 12.50 | 1.142 | 1.055 |
| 6.25 | 0.757 | 0.670 |
| 3.13 | 0.574 | 0.487 |
| 1.56 | 0.309 | 0.222 |
| 0.78 | 0.192 | 0.105 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-95% | 91% |
| EDTA plasma(n=5) | 88-103% | 96% |
| Heparin plasma(n=5) | 94-113% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-102% | 88-98% | 90-118% | 94-104% |
| EDTA plasma(n=5) | 83-93% | 81-111% | 83-111% | 93-106% |
| Heparin plasma(n=5) | 90-116% | 93-99% | 95-111% | 88-117% |