| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.405 | 2.316 |
| 10.00 | 1.582 | 1.493 |
| 5.00 | 1.023 | 0.934 |
| 2.50 | 0.771 | 0.682 |
| 1.25 | 0.570 | 0.481 |
| 0.63 | 0.367 | 0.278 |
| 0.31 | 0.323 | 0.234 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-118% | 106% |
| EDTA plasma(n=5) | 82-124% | 103% |
| Heparin plasma(n=5) | 91-109% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 80-119% | 85-98% | 82-101% | 87-113% |
| EDTA plasma(n=5) | 94-110% | 85-114% | 87-112% | 83-91% |
| Heparin plasma(n=5) | 93-120% | 84-110% | 94-107% | 88-98% |