| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.069 | 1.984 |
| 5.00 | 1.596 | 1.511 |
| 2.50 | 1.114 | 1.029 |
| 1.25 | 0.768 | 0.683 |
| 0.63 | 0.606 | 0.521 |
| 0.31 | 0.242 | 0.157 |
| 0.16 | 0.207 | 0.122 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-111% | 102% |
| EDTA plasma(n=5) | 90-128% | 109% |
| Heparin plasma(n=5) | 84-114% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-103% | 91-120% | 94-108% | 84-108% |
| EDTA plasma(n=5) | 81-115% | 82-119% | 82-95% | 85-108% |
| Heparin plasma(n=5) | 88-106% | 86-115% | 83-113% | 81-116% |