| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.101 | 2.027 |
| 25.00 | 1.668 | 1.594 |
| 12.50 | 1.224 | 1.150 |
| 6.25 | 1.004 | 0.930 |
| 3.13 | 0.519 | 0.445 |
| 1.56 | 0.310 | 0.236 |
| 0.78 | 0.157 | 0.083 |
| 0.00 | 0.074 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-120% | 101% |
| EDTA plasma(n=5) | 94-100% | 97% |
| Heparin plasma(n=5) | 85-107% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-111% | 91-100% | 84-113% | 86-91% |
| EDTA plasma(n=5) | 88-95% | 92-101% | 85-93% | 82-118% |
| Heparin plasma(n=5) | 82-102% | 91-103% | 82-110% | 80-97% |

關注微信公眾號