| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.935 | 1.856 |
| 10.00 | 1.547 | 1.468 |
| 5.00 | 1.082 | 1.003 |
| 2.50 | 0.817 | 0.738 |
| 1.25 | 0.601 | 0.522 |
| 0.63 | 0.303 | 0.224 |
| 0.31 | 0.280 | 0.201 |
| 0.00 | 0.079 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-117% | 104% |
| EDTA plasma(n=5) | 84-90% | 87% |
| Heparin plasma(n=5) | 80-94% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-107% | 80-113% | 87-112% | 81-105% |
| EDTA plasma(n=5) | 95-111% | 84-90% | 91-109% | 90-98% |
| Heparin plasma(n=5) | 83-91% | 87-109% | 83-119% | 82-119% |