| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 40.00 | 2.175 | 2.074 |
| 20.00 | 1.664 | 1.563 |
| 10.00 | 1.145 | 1.044 |
| 5.00 | 0.712 | 0.611 |
| 2.50 | 0.592 | 0.491 |
| 1.25 | 0.386 | 0.285 |
| 0.63 | 0.217 | 0.116 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-114% | 104% |
| EDTA plasma(n=5) | 88-126% | 107% |
| Heparin plasma(n=5) | 80-118% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-106% | 92-104% | 90-96% | 82-99% |
| EDTA plasma(n=5) | 89-95% | 86-116% | 90-117% | 84-112% |
| Heparin plasma(n=5) | 81-105% | 86-115% | 92-112% | 83-94% |