| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.158 | 2.062 |
| 500.00 | 1.614 | 1.518 |
| 250.00 | 1.124 | 1.028 |
| 125.00 | 0.831 | 0.735 |
| 62.50 | 0.607 | 0.511 |
| 31.25 | 0.322 | 0.226 |
| 15.63 | 0.248 | 0.152 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-97% | 91% |
| EDTA plasma(n=5) | 89-95% | 92% |
| Heparin plasma(n=5) | 86-102% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-113% | 89-102% | 82-100% | 89-103% |
| EDTA plasma(n=5) | 94-112% | 90-102% | 87-117% | 82-90% |
| Heparin plasma(n=5) | 85-93% | 83-119% | 86-111% | 91-113% |